Thermo Dionex AS-AP Autosampler P/N 074925
Thermo Dionex AS-AP Autosampler P/N 074925
Thermo Dionex AS-AP Autosampler P/N 074925
Thermo Dionex AS-AP Autosampler P/N 074925
Thermo Dionex AS-AP Autosampler P/N 074925
Thermo Dionex AS-AP Autosampler P/N 074925
Thermo Dionex AS-AP Autosampler P/N 074925
Thermo Dionex AS-AP Autosampler P/N 074925

Thermo Dionex AS-AP Autosampler P/N 074925

sku: 3370223
Out Of Stock
Ask a question
Condition : Excellent
Warranty : 30-Day Warranty, 100% Parts and Labor
Region : US
$5,500.00
Apply For Finance
Make Offer
Enquiry Here
ADD TO CART
  • Make Thermo
  • Hertz 50Hz/60Hz
  • Model Dionex AS-AP
  • Serial 22063363
  • Weight 57.0 Lb
  • Voltage 120V/220V
  • Includes Power cord included |
  • Warranty 30-Day Warranty, 100% Parts and Labor
  • Dimensions 26.0in x 23.0in x 20.0in
  • Shipping Type FedEx Ground
  • Item Condition Excellent
  • Manufacturing Date Does Not Apply

Product Information

This Thermo Dionex AS-AP Autosampler P/N 074925 is designed to provide high precision, reliability, and ruggedness while maintaining ease of use. It offers high-performance, automated sample processing for ion chromatography applications. The system has been tested for basic power functions, but it is missing sample trays. This autosampler is capable of simultaneous injection without cross contamination of chemistries, ensuring consistent operation at reproducible results. Equipped with various features such as a moving-needle design, sample preparation function, and well plate capabilities, it offers incredible flexibility for researchers and scientists. The Dionex AS-AP Autosampler can accommodate up to 81 10 mL vials and is suitable for a wide range of applications including trace analysis. Make: Thermo Model: Dionex AS-AP Condition: Used

Key Features

  • Simultaneous injection without cross contamination
  • Consistent operation with reproducible results
  • Moving-needle design for reliable sampling
  • High sample capacity and automated functions
  • Well plate capabilities
Specifications

  • Sample Capacity: 81 ea. 10 mL; 120 ea. 1.5 mL; Well Plates (3 96) or (3 384)
  • Injections per Vial: 199
  • Minimum Sample Volume: 10L can be sampled from a 300 L microvial; 20 L can be sampled from a 500 L microvial
  • Variable Volume Range: 1-1000L in 0.1 L increments; 100-7500 L in 1 L increments
  • Injection Precision: Fixed loop <0.3% RSD at 20 L; Partial loop <0.5% RSD at 20 L
  • Dilution Precision: <1.0% RSD for a 1:10 dilution
  • Dispensing Precision: <0.2% RSD (by weighing)
  • Carryover: <0.01% with 500 L flush volume
  • Operating Temperature Range: 460°C (40-140°F)
  • Power Requirements: 90-265 V ac, 47-63 Hz
Weight

Metric: 25.85 kg
Imperial: 57 lbs

Dimensions

Metric: 26.0 cm x 23.0 cm x 20.0 cm
Imperial: 10.24 in x 9.06 in x 7.87 in

Harmonized Code

90272000

Harmonized Code Details

90272000 - Instruments and apparatus for physical or chemical analysis

Testimonials

★★★★★

“REUZEit has been a great partner for our used equipment needs and always provide timely updates of new arrivals of consigned equipment.”

★★★★★

“Great company to work with. Tammy completed our first international equipment shipment with ease.”

★★★★★

“Fast response, open to adjusting schedule as needed, and great customer interaction.”

Need Answers? We’re Here to Help