Hewlett Packard G1362A Refractive Index Detector
- Make Hewlett Packard
- Hertz 50Hz/60Hz
- Model G1362A Refractive Index Detector
- Serial DE91600408
- Weight 48.0 Lb
- Voltage 120V/220V
- Includes Power cord included |
- Warranty 30-Day Warranty, 100% Parts and Labor
- Dimensions 25.0in x 22.0in x 15.0in
- Shipping Type FedEx Ground
- Item Condition Excellent
- Manufacturing Date Does Not Apply
Product Information
The Hewlett Packard 1100 HPLC G1362A Refractive Index Detector is designed for optical performance, GLP compliance, and easy maintenance. With advanced temperature controlled detector optics, the G1362A is ready to use within two hours of installation. This RID features uninterrupted operation from automatic solvent recycling and auto purge, along with integrated diagnostics for efficient troubleshooting. With built-in refractive index calibration and automatic light intensity control circuit, this RID ensures optimum performance. Easy access and maintenance to valves and capillaries through front. The G1362A acts as a universal detector, especially useful for substances that do not absorb visible light, such as lipids, polymers, and carbohydrates.
Key Features- Exhibits advanced temperature controlled detector optics for optimal performance
- Integrates automatic solvent recycling and auto purge for uninterrupted operation
- Includes built-in refractive index calibration and automatic light intensity control circuit
- Detection type: Refractive index
- Refractive index range: 1.00 - 1.75 RIU, calibrated
- Measurement range: 600 x 10-6 RIU
- Optical zeroing via set screw
- Optics temperature control: Ambient +5 to 55C
- Sample Cell: volume 8uL; maximum pressure 5 bar (0.5Mpa); maximum flow rate 5mL/minute
- Valves: Automatic purge and automatic solvent recycle
- Liquid contact materials: 316 stainless steel, teflon and quartz glass
- pH range: 2.3 - 9.5
- Short term noise: < 2.5 x 10-9 RIU
- Drift: < 200 x 10-9 RIU/hour
- Time programmable parameters: polarity, peak width
- Detector zero: automatic zero before analysis
- Analog outputs: Recorder/integrator - 100 mV or 1 V, output range selectable, one output
- Communications: Controller-area network (CAN), GPIB, RS-232C, LAN, APG
- GLP features: Early Maintenance Feedback (EMF), extensive diagnostics, error detection and display, electronic records of maintenance and errors, automated operational qualification/performance verification
Metric: 21.772 kg
Imperial: 48.0 lbs
Metric: 43.18 cm x 55.88 cm x 38.1 cm
Imperial: 17.0 in x 22.0 in x 15.0 in
902780
Harmonized Code Details902780 - Other instruments and apparatus using optical radiations (UV, visible, IR)
Testimonials