Thermo Scientific Diode Array Detector HL VH-D10-A
Thermo Scientific Diode Array Detector HL VH-D10-A
Thermo Scientific Diode Array Detector HL VH-D10-A
Thermo Scientific Diode Array Detector HL VH-D10-A
Thermo Scientific Diode Array Detector HL VH-D10-A
Thermo Scientific Diode Array Detector HL VH-D10-A
Thermo Scientific Diode Array Detector HL VH-D10-A
Thermo Scientific Diode Array Detector HL VH-D10-A
Thermo Scientific Diode Array Detector HL VH-D10-A
Thermo Scientific Diode Array Detector HL VH-D10-A

Thermo Scientific Diode Array Detector HL VH-D10-A

sku: 7000254
1 In Stock
Ask a question
Condition : P (Poor)
Warranty : AS-IS
Region : EU
$1,249.00$2,500.00
Apply For Finance
Make Offer
Enquiry Here
ADD TO CART
  • Make Thermo Scientific
  • Hertz 50Hz/60Hz
  • Model Diode Array Detector
  • Serial 8300304
  • Weight 23.0 Kg
  • Voltage 240V
  • Warranty AS-IS
  • Dimensions 96.0cm x 72.0cm x 46.0cm
  • Shipping Type FedEx Ground
  • Item Condition P (Poor)
  • Manufacturing Date 2014-04-30

Product Information

Thermo Scientific Vanquish Diode Array Detector HL VH-D10-A is an advanced detection device for HPLC and UHPLC systems. It offers excellent linearity and noise performance, allowing the detection of concentrated compounds and trace-level impurities. With LightPipe technology, it achieves superior signal-to-noise performance, making it an ideal choice for reliable and flexible detection in pharmaceutical and laboratory settings. This Thermo Scientific Diode Array Detector is known for its high sensitivity, light path, minimal dispersion, and GLP features. It is designed for UV-vis detection with up to 10 channels and offers an ultra-wide dynamic range.

Key Features

  • Ultra-wide dynamic range for detection of concentrated compounds and trace-level impurities
  • LightPipe technology for superior signal-to-noise performance
  • High sensitivity and long light path with minimal dispersion
  • GLP features for predictive performance functions and service monitoring
  • Up to 10 channels for UV-vis detection
Specifications

  • Detection Type: UV-vis Detection (Up to 10 Channels)
  • Detector Type: Single-Beam Diode Array Detector
  • Data Collection Rate: Up to 200 Hz
  • Drift: < 0.5 mAU/hour at 230nm
  • Flow Cell Volume: Standard, LightPipe flow cell, 10mm; High Sensitivity, Light Pipe flow cell, 60mm; LightPipe diagnostic cell
  • Lamp: Deuterium lamp
  • Linearity: < 5% at 2.0 AU (typically < 5% at 2.5 AU)
  • Max. Cell Operating Pressure: Limit 6 MPa (60 bar, 870 psi)
  • Noise: < 3 AU (at 230 nm)
  • No. of Channels: 10 + 3D field
  • PC Connection: USB 2.0; 3-port-HUB to connect further Vanquish modules
  • Photodiode Array: 1024 element
  • Spectral Bandwidth: Pixel resolution: 0.5nm (average); Optical resolution: 1nm (FWHM with 1nm slit width)
  • Temperature Range: 5 to 35°C; 20 to 80% RH (non-condensing)
  • Wavelength Range: 190 to 680 nm
  • Dimensions: 160 mm x 420 mm x 620 mm
  • Weight (Metric): 17 kg
  • Voltage: 100-240 V AC, 50/60 Hz, max. 150 W/255 VA
Weight

Metric: 17 kg
Imperial: 37.48 lb

Dimensions

Metric: 160 cm x 420 cm x 620 cm
Imperial: 63 in x 165.35 in x 244.09 in

Harmonized Code

902730

Harmonized Code Details

902730 - Spectrophotometers, spectrometers, and spectrographs using optical radiations (UV, visible, IR)

Testimonials

★★★★★

“REUZEit has been a great partner for our used equipment needs and always provide timely updates of new arrivals of consigned equipment.”

★★★★★

“Great company to work with. Tammy completed our first international equipment shipment with ease.”

★★★★★

“Fast response, open to adjusting schedule as needed, and great customer interaction.”

Need Answers? We’re Here to Help