Agilent G1315B DAD Diode Array Detector 1100 Series Precision Optics
Used, with a known minor cosmetic or mechanical defect, or a repair that may be needed. Still a valuable unit, but expect a modest investment to bring it back to full condition.
Agilent
Details
-
CategoryAnalytical
-
Dimension18.0in x 14.0in x 6.0in
-
SerialDE54627334
-
Shipping TypeFedEx Ground
-
Voltage / Hertz120V/220V / 50Hz/60Hz
Description
The Agilent 1100 Series G1315B Diode Array Detector, sourced from a working laboratory, features a flexible slit setting and offers maximum optical performance with its wide wavelength range of 190-950 nm. It is equipped with deuterium and tungsten lamps, and a 1024-element photodiode array. Despite being in fair condition with some minor issues like signal noise indicating possible detector electronics or lamp alignment problems, the device is high-value when operational. It has a slit width range of 1 to 16 nm for sensitivity and noise reduction, yet failed its self-test due to a signal noise value of 0.026mAU. The unit supports data evaluation via Agilent Chemstation for LC and operates on 120V/220V power at 50Hz/60Hz. This preowned unit includes a power cord but lacks data cables, a computer, or software.
Technician comment
Preowned unit sourced from a working lab; module powers on, lamp heats up, but front-panel error LED does not illuminate. Failed self-test due to signal noise value 0.026mAU indicating a likely detector electronics or lamp alignment issue. See photos for details & test results.
Key Features
- Flexible slit settings: 1 nm to 16 nm for optimal sensitivity
- Wavelength range: 190-950 nm
- Deuterium and tungsten lamps
- 1024-element photodiode array
- Operates on 120V/220V, 50Hz/60Hz
Specifications
- Wavelength range: 190-950 nm
- Light source: Deuterium and tungsten lamps
- Detection type: 1024-element photodiode array
- Slit width options: 1, 2, 4, 8, 16 nm
- Short term noise (ASTM): ± 1° x 10-5 AU at 254 and 750 nm
- Drift: 2° x 10-3 AU/hr at 254 nm
- Linear absorbance range: > 2 AU
- Wavelength accuracy: ± 1 nm
- Control software: Agilent Chemstation for LC
- Voltage: 120V/220V
- Frequency: 50Hz/60Hz
Weight
Imperial: 31.2 Pounds
Metric: 14.15 Kilograms
Shipping Dimensions
Imperial: 21.6 lb x 16.8 lb x 7.2 lb
Metric: 54.86 cm x 42.67 cm x 18.29 cm
Harmonized Code
9027.50.4050
Harmonized Code Details
9027.50.4050 - Instruments and apparatus for measuring or checking electricity quantities, for checking semiconductor wafers or devices; other. Suitable for laboratory use.